Maintenance
— 38
SPECTRO ARCOS FHX22 — 18.07.2019
6.6
Cleaning the load coil
Warning! Hot surface!
The Plasma torch, plasma interface, and load coil may be very hot.
Let the components cool for 10–15 minutes before commencing work on
them.
The load coil may corrode because of the sample vapors.
In order to clean the coil, proceed as follows:
1.
Switch off the instrument.
2.
Remove the Plasma torch, see page 32.
3.
Clean the load coil with a small brush (e.g. a toothbrush) and a grinding paste
(e.g. toothpaste).
Do not bend the load coil.
4.
Remove the grinding paste completely from the load coil.
5.
Reinstall the components.
The load coil is cleaned.
6.7
Optimizing the nebulizer gas flow
In order to optimize the process, proceed as follows:
1.
Call up the “Nebulizer Optimization” menu item via the instrument control
parameter dialog.
2.
Select a parameter set that suits the sample introduction system or solvent
used.
3.
Nebulize a solution that contains the specified elements in the defined
concentrations (see the online Help).
4.
Click the “Start” button in the dialog window.
The instrument optimizes the nebulizer gas flow.
Optimization of the nebulizer gas flow is now complete.
For additional information, refer to the online Help.