![Ametek SPECTRO ARCOS FHX22 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/ametek/spectro-arcos-fhx22/spectro-arcos-fhx22_original-operating-instructions_3044135019.webp)
Transportation/Setting-up
SPECTRO ARCOS FHX22 — 18.07.2019
— 19
4.4
Dust load
Avoiding contamination problems in the analysis requires a relatively dust-free
environment.
The maximum dust load should not exceed 36 million particles (0.5 µm or larger)
per cubic meter of air. Operating the instrument in an environment with a higher
dust load shortens maintenance intervals and may damage the instrument.
Example for comparison: A usual, clean office environment contains 18 to 36
million particles per cubic meter of indoor air.
To avoid instrument damage, the instrument must also be set up in an
environment which is free from corrosive vapors.
4.5
Storage
Storage conditions:
Unit must be stored in a dry place
Room temperature 2°C–40°C (35.6°F–104°F)
Avoid storing the instrument at very low temperatures.