![Ametek SPECTRO ARCOS FHX22 Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/ametek/spectro-arcos-fhx22/spectro-arcos-fhx22_original-operating-instructions_3044135032.webp)
Maintenance
— 32
SPECTRO ARCOS FHX22 — 18.07.2019
6.2
Cleaning the plasma torch
Warning! Hot surface!
The plasma torch, the plasma interface, and the induction coil can become
extremely.
Let the components cool for 10 to 15 minutes before commencing work on
them.
Warning! Sharp edges!
Glass parts may break when handled incorrectly and cause cutting injuries.
Proceed with due caution and/or take appropriate safety precautions (e.g.,
wear gloves).
Notice – Damage to instrument!
Plasma torch components that are not made of quartz glass may be damaged by
the cleaning solution or its vapors.
Clean only the quartz glass of the plasma torch in the cleaning solution.
To clean the plasma torch, proceed as follows:
1.
Switch off the plasma.
2.
Loosen and remove the clamp(s) between the plasma torch and the spray
chamber.
3.
Loosen the plasma torch by turning it counterclockwise.
4.
Remove the plasma torch.
5.
Disassemble the plasma torch.
Carefully remove the
adapter tube (6) from the
sleeve (5) by simultaneously
turning and pulling it.
Carefully remove the glass
element (1) from the holder
(2) by simultaneously turning
and pulling it.
Unscrew the sleeve (5) from
the holder (2).
Push/pull the injector tube
(3) from the holder (2).
Fig. 9