![Ametek SPECTRO ARCOS FHX22 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/ametek/spectro-arcos-fhx22/spectro-arcos-fhx22_original-operating-instructions_3044135020.webp)
Transportation/Setting-up
— 20
SPECTRO ARCOS FHX22 — 18.07.2019
4.6
Gas supply
Information! Measurement errors!
Impurities in the argon supply will produce measurement errors and/or will
make ICP operation impossible.
Ensure oil and grease-free installation of the argon supply line.
An argon supply system is required for the instrument.
A suitable pressure reducer is required for the argon supply system (bottles or
liquid gas tank).
Adjustable output pressure 0 to 10 bar
It is imperative not to exceed a maximum argon inlet pressure of 8 bar
(116 psi).
The tubing for the argon supply system is included in SPECTRO’s scope of supply
(5 m long, OD/ID = 8/6 mm). This tubing is screwed to the connection point on the
right-hand side of the instrument. The connection point to the instrument is
included in the scope of supply. The connector on-site is customer-specific and,
therefore, not part of the delivery.
The following flow quantities are used for aqueous samples during normal
operation:
Coolant:
9 to 20 l/min, typically 12 l/min
Auxiliary gas:
0.6 to 3 l/min, typically 0.8 l/min
Nebulizer gas:
0.4 to 1.5 l/min, typically 0.8 l/min