Maintenance
SPECTRO ARCOS FHX22 — 18.07.2019
— 37
9.
Clean the spray chamber.
First, use a suitable diluted acid solution (e.g., hydrochloric acid or nitric
acid 1:10).
Hazardous substance!
Observe the corresponding information on hazardous materials.
After this, use a laboratory cleaning agent for removing traces of fatty
substances (streaks) if required.
Now flush the spray chamber thoroughly with distilled water.
10. Reinstall the components.
11. Optimize the nebulizer gas flow, see page 38.
Cleaning of the spray chamber is now complete.
6.5
Checking the aerosol formation of the nebulizer
In order to proceed with the inspection, follow the steps below:
1.
Remove the nebulizer, see page 35, but without disconnecting the argon
connection.
2.
Call up the “Solid State Generator” menu item via the instrument control
parameter dialog.
3.
Adjust the nebulizer gas flow determined for the nebulizer.
4.
Using a spray bottle, force
distilled water through the
nebulizer connection for sample
introduction.
The spray coming out must
be evenly V-shaped. If this
is not the case, the
nebulizer may be clogged
and must be cleaned, see
page 35.
Fig. 13
5.
Reinstall the components.
Checking of aerosol generation is now complete.