Maintenance
— 36
SPECTRO ARCOS FHX22 — 18.07.2019
10. Then flush the nebulizer with distilled water and let it dry completely.
11. Reinstall the components.
12. Check aerosol generation and optimize the gas flow of the nebulizer, see
Cleaning of the nebulizer is now complete.
6.4
Cleaning the spray chamber (standard Scott Chamber)
Warning! Risk of injury!
There may be acid residues or other hazardous residues from sample material in
the nebulizer and the spray chamber.
Observe the corresponding information on hazardous materials.
Warning! Sharp edges!
Glass parts may break when handled incorrectly and cause cutting injuries.
Proceed with due caution and/or take appropriate safety precautions (e.g.,
wear gloves).
To clean the spray chamber, proceed as follows:
1.
Flush the sample introduction system for approx. 10 minutes with a suitable
blank solution, e.g., distilled water for aqueous applications.
2.
Switch off the plasma.
3.
Loosen the Ar connection of the nebulizer.
4.
Loosen the clamp between the Plasma torch and the spray chamber.
5.
Carefully pull off the spray chamber.
6.
Remove the spray chamber.
7.
Separate the nebulizer from the spray chamber by slightly twisting it.
8.
Disconnect the waste water connection from the drainage of the spray
chamber and remove the cap from the spray chamber by slightly turning it.