Troubleshooting
— 52
SPECTRO ARCOS FHX22 — 18.07.2019
Malfunction
Cause
Remedy
Error message:
The plasma is
extinguished.
Analytical load too high.
Check whether the adjustable
parameters harmonize with the
selected method.
Error message:
Plasma does not ignite.
Plasma torch is
contaminated.
Coil geometry faulty.
Check the plasma torch; clean if
necessary.
Check the position and the
geometry of the coil, align if
necessary.
Error message:
Torch position undefined.
Failure to reach the
selected plasma torch
position. Only for
MultiView option.
Check whether the Plasma torch
position has been reached.
Error message:
MultiView blind plug
missing.
Bling plug not inserted or
not connected.
Check the position of the blind
plug and monitoring cable
connector.
Plasma torch is melting.
Plasma torch is incorrectly
positioned inside the
induction coil.
Check whether the plasma torch
is positioned correctly.
Aerosol tube clogged.
Clean plasma torch
Leak in argon connection
to the plasma torch.
Check the O-rings for sealing the
plasma torch and replace them, if
necessary.
Auxiliary gas flow set too
low.
Increase auxiliary gas flow.
Leak in nebulizer system.
Localize leak using soapy
solution and seal the leak.
Plasma cannot be
ignited.
No Tesla coil spark.
Fault with insulation for torch
holder (service)
Plasma torch not correctly
centered within the load
coil.
Check the load coil setting.
Oxygen in sample
introduction system.
Purge the sample introduction
system with argon
Wrong argon quality.
Argon quality of 4.5 or higher is
required (purity of 99.995 % or
higher).
Asymmetrical plasma.
Argon flushing from the
OPI/SPI too high.
Reduce the flow rate.
Plasma chamber
extraction set too high.
Reduce the extraction capacity.