:TRIGger:CAN:SIGNal:DEFinition
(see
Command Syntax
:TRIGger:CAN:SIGNal:DEFinition <value>
<value> ::= {CANH | CANL | RX | TX | DIFFerential}
The :TRIGger:CAN:SIGNal:DEFinition command sets the CAN signal type
when :TRIGger:CAN:TRIGger is set to SOF (start of frame). These signals
can be set to:
Dominant high signal:
•
CANH — the actual CAN_H differential bus signal.
Dominant low signals:
•
CANL — the actual CAN_L differential bus signal.
•
RX — the Receive signal from the CAN bus transceiver.
•
TX — the Transmit signal to the CAN bus transceiver.
•
DIFFerential — the CAN differential bus signal connected to an analog
source channel using a differential probe.
Query Syntax
:TRIGger:CAN:SIGNal:DEFinition?
The :TRIGger:CAN:SIGNal:DEFinition? query returns the current CAN
signal type.
Return Format
<value><NL>
<value> ::= DIFF
See Also
•
"Introduction to :TRIGger Commands"
•
•
":TRIGger:CAN:SIGNal:BAUDrate"
•
•
N O T E
With 6000 Series oscilloscope software version 3.50 or greater, this command is available,
but the only legal value is DIFF.
Содержание DSO6014A/L
Страница 1: ...A Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 16: ...16 Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 30: ...30 Agilent 6000 Series Oscilloscopes Programmer s Reference 1 What s New...
Страница 540: ...540 Agilent 6000 Series Oscilloscopes Programmer s Reference 6 Error Messages...
Страница 562: ...562 Agilent 6000 Series Oscilloscopes Programmer s Reference 7 Status Reporting...
Страница 584: ...584 Agilent 6000 Series Oscilloscopes Programmer s Reference 8 More About Oscilloscope Commands...
Страница 624: ...624 Agilent 6000 Series Oscilloscopes Programmer s Reference 9 Programming Examples...