Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
287
:POD Commands
Control all oscilloscope functions associated with groups of digital
channels. See
"Introduction to :POD<n> Commands"
Introduction to
:POD<n>
Commands
<n> ::= {1 | 2}
The POD subsystem commands control the viewing and threshold of
groups of digital channels.
POD1 ::= D0- D7
POD2 ::= D8- D15
Reporting the Setup
Use :POD1? or :POD2? to query setup information for the POD subsystem.
Return Format
The following is a sample response from the :POD1? query. In this case,
the query was issued following a *RST command.
:POD1:DISP 0;THR +1.40E+00
Table 58 :POD<n> Commands Summary
Command
Query
Options and Query Returns
:POD<n>:DISPlay {{0 |
OFF} | {1 | ON}} (see
:POD<n>:DISPlay? (see
{0 | 1}
<n> ::= 1-2 in NR1 format
:POD<n>:SIZE <value>
(see
:POD<n>:SIZE? (see
<value> ::= {SMALl | MEDium | LARGe}
:POD<n>:THReshold
<type>[suffix] (see
:POD<n>:THReshold? (see
<n> ::= 1-2 in NR1 format
<type> ::= {CMOS | ECL | TTL | <user
defined value>}
<user defined value> ::= value in
NR3 format
[suffix] ::= {V | mV | uV }
N O T E
These commands are only valid for the MSO models.
Содержание DSO6014A/L
Страница 1: ...A Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 16: ...16 Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 30: ...30 Agilent 6000 Series Oscilloscopes Programmer s Reference 1 What s New...
Страница 540: ...540 Agilent 6000 Series Oscilloscopes Programmer s Reference 6 Error Messages...
Страница 562: ...562 Agilent 6000 Series Oscilloscopes Programmer s Reference 7 Status Reporting...
Страница 584: ...584 Agilent 6000 Series Oscilloscopes Programmer s Reference 8 More About Oscilloscope Commands...
Страница 624: ...624 Agilent 6000 Series Oscilloscopes Programmer s Reference 9 Programming Examples...