Agilent 6000 Series Oscilloscopes Programmer's Reference
3
Commands by Subsystem
:CHANnel<n>:PROBe:STYPe
(see
Command Syntax
:CHANnel<n>:PROBe:STYPe <signal type>
<signal type> ::= {DIFFerential | SINGle}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :CHANnel<n>:PROBe:STYPe command sets the channel probe signal
type (STYPe) to differential or single- ended when using the 113xA Series
probes and determines how offset is applied.
When single- ended is selected, the :CHANnel<n>:OFFset command changes
the offset value of the probe amplifier. When differential is selected, the
:CHANnel<n>:OFFset command changes the offset value of the channel
amplifier.
Query Syntax
:CHANnel<n>:PROBe:STYPe?
The :CHANnel<n>:PROBe:STYPe? query returns the current probe signal
type setting for the selected channel.
Return Format
<signal type><NL>
<signal type> ::= {DIFF | SING}
See Also
•
"Introduction to :CHANnel<n> Commands"
•
N O T E
This command is valid only for the 113xA Series probes.
Содержание DSO6014A/L
Страница 1: ...A Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 16: ...16 Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 30: ...30 Agilent 6000 Series Oscilloscopes Programmer s Reference 1 What s New...
Страница 540: ...540 Agilent 6000 Series Oscilloscopes Programmer s Reference 6 Error Messages...
Страница 562: ...562 Agilent 6000 Series Oscilloscopes Programmer s Reference 7 Status Reporting...
Страница 584: ...584 Agilent 6000 Series Oscilloscopes Programmer s Reference 8 More About Oscilloscope Commands...
Страница 624: ...624 Agilent 6000 Series Oscilloscopes Programmer s Reference 9 Programming Examples...