402
Agilent 6000 Series Oscilloscopes Programmer's Reference
3
Commands by Subsystem
:TRIGger:LIN:SAMPlepoint
(see
Command Syntax
:TRIGger:LIN:SAMPlepoint <value>
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
The :TRIGger:LIN:SAMPlepoint command sets the point during the bit time
where the bit level is sampled to determine whether the bit is dominant or
recessive. The sample point represents the percentage of time between the
beginning of the bit time to the end of the bit time.
Query Syntax
:TRIGger:LIN:SAMPlepoint?
The :TRIGger:LIN:SAMPlepoint? query returns the current LIN sample
point setting.
Return Format
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
See Also
•
"Introduction to :TRIGger Commands"
•
•
N O T E
The sample point values are not limited by the baud rate.