:TRIGger:DURation:RANGe
(see
Command Syntax
:TRIGger:DURation:RANGe <greater than time>[<suffix>],
<less than time>[<suffix>]
<greater than time> ::= minimum duration in seconds
(10 ns - 9.99 seconds) in NR3 format
<less than time> ::= maximum duration in seconds
(15 ns - 10 seconds) in NR3 format
<suffix> ::= {s | ms | us | ns | ps}
The :TRIGger:DURation:RANGe command sets the duration for the defined
pattern when the :TRIGger:DURation:QUALifier command is set to INRange
or OUTRange.
Query Syntax
:TRIGger:DURation:RANGe?
The :TRIGger:DURation:RANGe? query returns the duration time for the
defined pattern.
Return Format
<greater than time>,<less than time><NL>