:DIGital<n>:SIZE
(see
Command Syntax
:DIGital<n>:SIZE <value>
<n> ::= An integer, 0, 1,..,15, is attached as a suffix to the command
and defines the logic channel that is affected by the command.
<value> ::= {SMALl | MEDium | LARGe}
The :DIGital<n>:SIZE command specifies the size of digital channels on the
display. Sizes are set for all digital channels. Therefore, if you set the size
on digital channel 0 (for example), the same size is set on channels 1
through 15 as well.
Query Syntax
:DIGital<n>:SIZE?
The :DIGital<n>:SIZE? query returns the size setting for the specified
digital channels.
Return Format
<size_value><NL>
<size_value> ::= {SMAL | MED | LARG}
See Also
•
"Introduction to :DIGital<n> Commands"
•
•
N O T E
This command is only valid for the MSO models.