Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
397
:TRIGger:IIC:TRIGger:QUALifier
(see
Command Syntax
:TRIGger:IIC:TRIGger:QUALifier <value>
<value> ::= {EQUal | NOTequal | LESSthan | GREaterthan}
The :TRIGger:IIC:TRIGger:QUALifier command sets the IIC data qualifier
when TRIGger:IIC:TRIGger[:TYPE] is set to READEeprom.
Query Syntax
:TRIGger:IIC:TRIGger:QUALifier?
The :TRIGger:IIC:TRIGger:QUALifier? query returns the current IIC data
qualifier value.
Return Format
<value><NL>
<value> ::= {EQUal | NOTequal | LESSthan | GREaterthan}
See Also
•
"Introduction to :TRIGger Commands"
•
•