Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
205
:EXTernal:IMPedance
(see
Command Syntax
:EXTernal:IMPedance <value>
<value> ::= {ONEMeg | FIFTy}
The :EXTernal:IMPedance command selects the input impedance setting for
the external trigger. The legal values for this command are ONEMeg
(1 M
Ω
) and FIFTy (50
Ω
).
Query Syntax
:EXTernal:IMPedance?
The :EXTernal:IMPedance? query returns the current input impedance
setting for the external trigger.
Return Format
<impedance value><NL>
<impedance value> ::= {ONEM | FIFT}
See Also
•
"Introduction to :EXTernal Trigger Commands"
•
"Introduction to :TRIGger Commands"
•
N O T E
You can set external trigger input impedance to FIFTy (50
Ω
) on the 2-channel, 300 MHz,
500 MHz, and 1 GHz bandwidth oscilloscope models.