Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
355
:TRIGger:DURation:GREaterthan
(see
Command Syntax
:TRIGger:DURation:GREaterthan <greater than time>[<suffix>]
<greater than time> ::= minimum trigger duration in seconds
(5 ns - 10 seconds) in NR3 format
<suffix> ::= {s | ms | us | ns | ps }
The :TRIGger:DURation:GREaterthan command sets the minimum duration
for the defined pattern when :TRIGger:DURation:QUALifier is set to
GREaterthan. The command also sets the timeout value when the
:TRIGger:DURation:QUALifier is set to TIMeout.
Query Syntax
:TRIGger:DURation:GREaterthan?
The :TRIGger:DURation:GREaterthan? query returns the minimum duration
time for the defined pattern.
Return Format
<greater than time><NL>