Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
159
:CALibrate Commands
Utility commands for viewing calibration status and for starting the user
calibration procedure. See
"Introduction to :CALibrate Commands"
Introduction to
:CALibrate
Commands
The CALibrate subsystem provides utility commands for:
•
Determining the state of the calibration factor protection switch
(CAL PROTECT).
•
Saving and querying the calibration label string.
•
Reporting the calibration time and date.
•
Reporting changes in the temperature since the last calibration.
•
Starting the user calibration procedure.
Table 49 :CALibrate Commands Summary
Command
Query
Options and Query Returns
n/a
:CALibrate:DATE? (see
<return value> ::=
<day>,<month>,<year>; all in NR1
format
:CALibrate:LABel
<string> (see
)
:CALibrate:LABel? (see
<string> ::= quoted ASCII string up
to 32 characters
:CALibrate:STARt (see
n/a
n/a
n/a
:CALibrate:STATus? (see
<return value> ::=
ALL,<status_code>,<status_string>
<status_code> ::= an integer status
code
<status_string> ::= an ASCII status
string
n/a
:CALibrate:SWITch? (see
{PROTected | UNPRotected}
n/a
:CALibrate:TEMPerature?
(see
<return value> ::= degrees C delta
since last cal in NR3 format
n/a
:CALibrate:TIME? (see
<return value> ::=
<hours>,<minutes>,<seconds>; all in
NR1 format
Содержание DSO6014A/L
Страница 1: ...A Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 16: ...16 Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 30: ...30 Agilent 6000 Series Oscilloscopes Programmer s Reference 1 What s New...
Страница 540: ...540 Agilent 6000 Series Oscilloscopes Programmer s Reference 6 Error Messages...
Страница 562: ...562 Agilent 6000 Series Oscilloscopes Programmer s Reference 7 Status Reporting...
Страница 584: ...584 Agilent 6000 Series Oscilloscopes Programmer s Reference 8 More About Oscilloscope Commands...
Страница 624: ...624 Agilent 6000 Series Oscilloscopes Programmer s Reference 9 Programming Examples...