Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
283
:MEASure:VTIMe
(see
Query Syntax
:MEASure:VTIMe? <vtime_argument>[,<source>]
<vtime_argument> ::= time from trigger in seconds
<source> ::= {<digital channels> | CHANnel<n> | FUNCtion | MATH}
<digital channels> ::= DIGital0,..,DIGital15 for the MSO models
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:VTIMe? query returns the value at a specified time on the
source specified with :MEASure:SOURce. The specified time must be on
the screen and is referenced to the trigger event. If the optional source
parameter is specified, the current source is modified.
Return Format
<value><NL>
<value> ::= value at the specified time in NR3 format
See Also
•
"Introduction to :MEASure Commands"
•
•
•
N O T E
This query is not available if the source is FFT (Fast Fourier Transform).
Содержание DSO6014A/L
Страница 1: ...A Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 16: ...16 Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 30: ...30 Agilent 6000 Series Oscilloscopes Programmer s Reference 1 What s New...
Страница 540: ...540 Agilent 6000 Series Oscilloscopes Programmer s Reference 6 Error Messages...
Страница 562: ...562 Agilent 6000 Series Oscilloscopes Programmer s Reference 7 Status Reporting...
Страница 584: ...584 Agilent 6000 Series Oscilloscopes Programmer s Reference 8 More About Oscilloscope Commands...
Страница 624: ...624 Agilent 6000 Series Oscilloscopes Programmer s Reference 9 Programming Examples...