Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
245
:MEASure:PERiod
[<source>] (see
:MEASure:PERiod?
[<source>] (see
<source> ::= {CHANnel<n> | FUNCtion
| MATH} for DSO models
<source> ::= {CHANnel<n> |
DIGital0,..,DIGital15 | FUNCtion |
MATH} for MSO models
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= waveform period
in seconds in NR3 format
:MEASure:PHASe
[<source1>]
[,<source2>] (see
:MEASure:PHASe?
[<source1>]
[,<source2>] (see
<source1,2> ::= {CHANnel<n> |
FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= the phase angle
value in degrees in NR3 format
:MEASure:PREShoot
[<source>] (see
:MEASure:PREShoot?
[<source>] (see
<source> ::= {CHANnel<n> | FUNCtion
| MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= the percent of
preshoot of the selected waveform in
NR3 format
:MEASure:PWIDth
[<source>] (see
:MEASure:PWIDth?
[<source>] (see
<source> ::= {CHANnel<n> | FUNCtion
| MATH} for DSO models
<source> ::= {CHANnel<n> |
DIGital0,..,DIGital15 | FUNCtion |
MATH} for MSO models
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= width of positive
pulse in seconds in NR3 format
:MEASure:RISEtime
[<source>] (see
:MEASure:RISEtime?
[<source>] (see
<source> ::= {CHANnel<n> | FUNCtion
| MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= rise time in
seconds in NR3 format
:MEASure:SDEViation
[<source>] (see
:MEASure:SDEViation?
[<source>] (see
<source> ::= {CHANnel<n> | FUNCtion
| MATH}
<n> ::= 1-2 or 1-4 in NR1 format
<return_value> ::= calculated std
deviation in NR3 format
Table 57 :MEASure Commands Summary (continued)
Command
Query
Options and Query Returns
Содержание DSO6014A/L
Страница 1: ...A Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 16: ...16 Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 30: ...30 Agilent 6000 Series Oscilloscopes Programmer s Reference 1 What s New...
Страница 540: ...540 Agilent 6000 Series Oscilloscopes Programmer s Reference 6 Error Messages...
Страница 562: ...562 Agilent 6000 Series Oscilloscopes Programmer s Reference 7 Status Reporting...
Страница 584: ...584 Agilent 6000 Series Oscilloscopes Programmer s Reference 8 More About Oscilloscope Commands...
Страница 624: ...624 Agilent 6000 Series Oscilloscopes Programmer s Reference 9 Programming Examples...