Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
215
:FUNCtion:DISPlay
(see
Command Syntax
:FUNCtion:DISPlay <display>
<display> ::= {{1 | ON} | {0 | OFF}}
The :FUNCtion:DISPlay command turns the display of the function on or
off. When ON is selected, the function performs as specified using the
other FUNCtion commands. When OFF is selected, function is neither
calculated nor displayed.
Query Syntax
:FUNCtion:DISPlay?
The :FUNCtion:DISPlay? query returns whether the function display is on
or off.
Return Format
<display><NL>
<display> ::= {1 | 0}