Chapter 4 Features and Functions
Scanning
4
4
Idle State
The instrument is considered to be in the idle state whenever it is not
operating within any one of the layers. When the instrument is taken out
of the idle state, the SCAN annunciator lights up and operation proceeds
to the arm layer. From the remote interface, the
INITiate
command
takes the instrument out of the idle state.
Arm Layer
The instrument requires an arm source to allow operation to proceed into
the trigger layer. You can select the arm source from the following:
•
TIMER — with timer arm source selected, the instrument will not
proceed to the trigger layer unless the specified time interval has
elapsed.
•
IMM — with IMM (default) arm source selected, operation
immediately proceeds to the trigger layer as soon as the instrument is
taken out of the idle state.
•
BUS — with bus arm source selected, the instrument will not proceed
to the trigger layer unless a
GET
or a
*TRG
command is received, or
Step on the front-panel is pressed.
•
EXT — with external (EXT) arm source selected, the instrument will
not proceed to the trigger layer unless an external trigger is received
from the specified trigger-in line.
•
MIX — with mix arm source selected, the instrument will not proceed
to the trigger layer unless a BUS event or EXTernal event occurs.
•
HOLD — with hold arm source selected, the instrument will wait for
a
TRIGger:IMMediate
command before proceeding.
Содержание Agilent 3499A
Страница 13: ...1 1 Quick Start ...
Страница 27: ...2 2 Front Panel Operation ...
Страница 55: ...3 3 System Overview ...
Страница 77: ...4 4 Features and Functions ...
Страница 113: ...5 5 Remote Interface Reference ...
Страница 164: ...164 5 ...
Страница 165: ...6 6 Error Messages ...
Страница 173: ...7 7 Plug in Modules ...
Страница 256: ...256 Chapter 7 Plug in Modules 44475A Breadboard Module 7 ...
Страница 261: ...261 Chapter 7 Plug in Modules 44476B Microwave Switch Module 4 7 The 44476B is shown below ...
Страница 268: ...268 Chapter 7 Plug in Modules 44478A B 1 3 GHz Dual 4 to 1 MUX Modules 7 The 44478A B is shown below ...
Страница 286: ...286 Chapter 7 Plug in Modules Terminals and Connections Information 7 ...
Страница 288: ...288 7 ...
Страница 289: ...8 8 Application Programs ...
Страница 299: ...9 9 Specifications ...
Страница 343: ...343 Chapter 9 Specifications 44475A Breadboard Module 4 9 ...