163
Chapter 5 Remote Interface Reference
About the SCPI Language
4
5
Parameters
These are parameter types used with the SCPI language:
Parameters shown within square brackets ([ ]) are optional. (Note that
the brackets are not part of the command and are not sent to the
instrument.) If you do not specify a value for an optional parameter, the
instrument uses the default value. For example, consider the
ARM:COUNt?[<MIN|MAX>]
command. If you send the command without
specifying a parameter, the present
ARM:COUNt
setting is returned. If
you send the
MIN
parameter, the command returns the minimum count
available. If you send the
MAX
parameter, the command returns the
maximum count available. Be sure to place a space between the
command and the parameter.
Linking Commands
To link IEEE 488.2 Common Commands with SCPI Commands, use a
semicolon between the commands. For example:
*RST;CONF:EXT:OUTP ON
or
TRIG:SOUR HOLD;*TRG
To link multiple SCPI commands, use both a semicolon and a colon
between the commands. For example:
ARM:COUN 1;:TRIG:SOUR EXT
Parameter Type
Description
Numeric
Accepts all commonly used decimal representations of number including
optional signs, decimal points, and scientific notation.
123, 123E2, -123, -1.23E2, .123, 1.23E-2, 1.23000E-01.
Special cases include MINimum, MAXimum, and DEFault.
Boolean
Represents a single binary condition that is either true or false.
ON, OFF, 1, 0
Discrete
Selects from a finite number of values. These parameters use mnemonics to
represent each valid setting.
An example is the TRIGger:SOURce <source> command
where source can be BUS, EXT, HOLD, or IMM.
Содержание Agilent 3499A
Страница 13: ...1 1 Quick Start ...
Страница 27: ...2 2 Front Panel Operation ...
Страница 55: ...3 3 System Overview ...
Страница 77: ...4 4 Features and Functions ...
Страница 113: ...5 5 Remote Interface Reference ...
Страница 164: ...164 5 ...
Страница 165: ...6 6 Error Messages ...
Страница 173: ...7 7 Plug in Modules ...
Страница 256: ...256 Chapter 7 Plug in Modules 44475A Breadboard Module 7 ...
Страница 261: ...261 Chapter 7 Plug in Modules 44476B Microwave Switch Module 4 7 The 44476B is shown below ...
Страница 268: ...268 Chapter 7 Plug in Modules 44478A B 1 3 GHz Dual 4 to 1 MUX Modules 7 The 44478A B is shown below ...
Страница 286: ...286 Chapter 7 Plug in Modules Terminals and Connections Information 7 ...
Страница 288: ...288 7 ...
Страница 289: ...8 8 Application Programs ...
Страница 299: ...9 9 Specifications ...
Страница 343: ...343 Chapter 9 Specifications 44475A Breadboard Module 4 9 ...