Chapter 6 Error Messages
Execution Errors
4
6
-161
Invalid block data
For a definite-length block, the number of types of data sent does not
match the number of bytes that you specified in the block header
-168
Block data not allowed
Data was sent to the instrument in SCPI definite length block format but
this command does not accept this format.
-178
Expression data not allowed
A channel list was received but is not allowed for this command.
Example:
SYST:CTYPE? (@100)
-222
Data out of range
A numeric parameter value is outside the valid range for this command.
Example:
ARM:COUNT -3
-223
Too much data
A character string was received but could not be executed because the
string length was more than 13 characters. This error can be generated
by the
DIAGnostic:DISPlay
command.
-224
Illegal parameter value
A discrete parameter was received which was not a valid choice for this
command. You may have used an invalid parameter choice.
Example:
TRIG:SOURCE ALARM
(ALARM is not a valid choice)
-310
System error
A firmware defect has been found. This is not a fatal error but you should
contact your nearest Agilent Technologies Service Center (see page 8) if
this error is reported.
-350
Queue overflow
The error queue is full because more than 10 errors have occurred. No
additional errors are stored until you remove errors from the queue. The
error queue is cleared by the
*CLS
(clear status) command or when power
is cycled. The errors are also cleared when you read the queue.
-410
Query INTERRUPTED
A command was received which sends data to the output buffer, but the
output buffer contained data from a previous command (the previous data
is not overwritten). The output buffer is cleared when power has been off
or after a bus Device Clear.
Содержание Agilent 3499A
Страница 13: ...1 1 Quick Start ...
Страница 27: ...2 2 Front Panel Operation ...
Страница 55: ...3 3 System Overview ...
Страница 77: ...4 4 Features and Functions ...
Страница 113: ...5 5 Remote Interface Reference ...
Страница 164: ...164 5 ...
Страница 165: ...6 6 Error Messages ...
Страница 173: ...7 7 Plug in Modules ...
Страница 256: ...256 Chapter 7 Plug in Modules 44475A Breadboard Module 7 ...
Страница 261: ...261 Chapter 7 Plug in Modules 44476B Microwave Switch Module 4 7 The 44476B is shown below ...
Страница 268: ...268 Chapter 7 Plug in Modules 44478A B 1 3 GHz Dual 4 to 1 MUX Modules 7 The 44478A B is shown below ...
Страница 286: ...286 Chapter 7 Plug in Modules Terminals and Connections Information 7 ...
Страница 288: ...288 7 ...
Страница 289: ...8 8 Application Programs ...
Страница 299: ...9 9 Specifications ...
Страница 343: ...343 Chapter 9 Specifications 44475A Breadboard Module 4 9 ...