106
Chapter 4 Features and Functions
State Storage
4
State Storage
The 3499A/B/C provides the capability to store setups, and then recall
them to put the instrument back into a known configuration.
Information that can be stored includes:
•
Status of all relay channels (open or closed)
•
Status of digital I/O ports (input or output, flow control, etc.)
•
Special module configurations (2-wire, card pair, etc.)
•
Scan lists and configuration (arm source, trigger source, etc.)
•
Remote interface settings (address, baud, etc.)
Additionally, you can include a stored state in a scan list as a channel.
Firmware Revision Differences
State storage behavior differs depending upon the firmware revision of
your 3499A/B/C (see page 59).
•
Firmware revision 1.0 must be upgraded. Please contact your nearest
Agilent Technologies Office for details.
•
Firmware revisions 2.0 and 3.0 allow up to 10 states to be stored.
Firmware revision 4.0 allows up to 50 stored states.
•
Stored states are lost if power is removed for firmware revisions 2.0
and 3.0. Stored states are preserved when power is removed for
firmware revision 4.0.
•
Firmware revision 4.0 allows you store a state and apply that state
when power is applied to the instrument.
•
Storage memory locations are numbered 01 to 10 for Firmware
revision 2.0 and 3.0. Memory locations are number 01 to 50 for
Firmware revision 4.0.
•
Before recalling a stored setup, the instrument verifies that all
module types and slot assignment match the setup. If a mismatch is
detected, an error (“RECALL FAILED”) will occur.
Содержание Agilent 3499A
Страница 13: ...1 1 Quick Start ...
Страница 27: ...2 2 Front Panel Operation ...
Страница 55: ...3 3 System Overview ...
Страница 77: ...4 4 Features and Functions ...
Страница 113: ...5 5 Remote Interface Reference ...
Страница 164: ...164 5 ...
Страница 165: ...6 6 Error Messages ...
Страница 173: ...7 7 Plug in Modules ...
Страница 256: ...256 Chapter 7 Plug in Modules 44475A Breadboard Module 7 ...
Страница 261: ...261 Chapter 7 Plug in Modules 44476B Microwave Switch Module 4 7 The 44476B is shown below ...
Страница 268: ...268 Chapter 7 Plug in Modules 44478A B 1 3 GHz Dual 4 to 1 MUX Modules 7 The 44478A B is shown below ...
Страница 286: ...286 Chapter 7 Plug in Modules Terminals and Connections Information 7 ...
Страница 288: ...288 7 ...
Страница 289: ...8 8 Application Programs ...
Страница 299: ...9 9 Specifications ...
Страница 343: ...343 Chapter 9 Specifications 44475A Breadboard Module 4 9 ...