![Agilent Technologies 8935 Series Скачать руководство пользователя страница 85](http://html1.mh-extra.com/html/agilent-technologies/8935-series/8935-series_manual_2868625085.webp)
84
Chapter 3
Troubleshooting
Manual Troubleshooting Procedures (Step 4)
3. Set the RF signal generator’s frequency to 100, then 500, 900, and
1800 MHz. For each frequency reset the
Tune Freq
to that
frequency. The Test Set’s measurements should read as follows:
a.
TX Power
should read approximately 0.001 W for each frequency.
b.
Frequency
should read 100, 500, 900, and 1800 MHz respectively.
c. Press the
Spec Anl
key to access the analog spectrum analyzer.
Observe the level and frequency of the signal.
Isolating the RF Source
The RF Generator function uses the following assemblies. Refer to
Figure 3-19
and the block diagrams in
Chapter 10 , “Block Diagrams,”
on page 283
.
• A2A120 LO IF/IQ Modulator
• A2A25 Signal Generator Synthesizer
• A2A24 Output Section
• A2A110 Upconverter
Figure 3-19
Isolating the RF Source
OPP
100 W
Attenuator
Splitter
Output
ANT
IN
RF
IN/OUT
DUPLEX
OUT
Input
Attenuator
Power
Detector
Up
Output
Section
IQ
Modulator
Sig Gen
Synthesizer
RF Generator
Reference
RF I/O
RPP
Attenuator
Converter
TO RF ANALYZER
DC Block
Содержание 8935 Series
Страница 34: ...34 Chapter 1 General Information Trademark Acknowledgements...
Страница 35: ...35 2 Product Information This chapter contains general information about the Test Set and how to service it...
Страница 93: ...92 Chapter 3 Troubleshooting Product Verification...
Страница 102: ...101 5 Disassembly This chapter explains how to disassemble the Test Set for major assembly replacement...
Страница 115: ...114 Chapter 5 Disassembly A2 Disassembly Figure 5 9 PC Board Assemblies...
Страница 135: ...134 Chapter 5 Disassembly Wire Cable Information...
Страница 139: ...138 Chapter 6 Replaceable Parts Parts Identification Covers and Chassis Parts Figure 6 2 External and Internal Covers...
Страница 150: ...Chapter 6 149 Replaceable Parts Parts Identification Figure 6 14 Cables Panel and Bottom Side Views...
Страница 157: ...156 Chapter 6 Replaceable Parts Parts List...
Страница 158: ...157 7 Periodic Adjustments This chapter contains the periodic adjustment procedures for the Test Set...
Страница 173: ...172 Chapter 7 Periodic Adjustments Eb No Calibration Program Description...
Страница 233: ...232 Chapter 8 Performance Tests CDMA Analyzer Modulation Accuracy Performance Test 29...
Страница 283: ...282 Chapter 9 Performance Test Records CDMA Analyzer Modulation Accuracy Performance Test 29 Record...
Страница 321: ...320 Chapter 10 Block Diagrams Instrument Control Section...
Страница 322: ...321 A Error Messages...
Страница 337: ...336 Appendix A Error Messages Non Recoverable Firmware Error...