Chapter 7
167
Periodic Adjustments
Periodic Calibration Menu Descriptions
Figure 7-4
Measurement (A2A33) Assembly Test Points
Audio Frequency Generator Gain
The gain of the following paths is calibrated:
• The internal paths that run from Audio Frequency Generators 1 and
2 (individually) through the Modulation Distribution assembly, to
the monitor select output, then onto Audio Analyzer 1 to the DVM.
• The paths that run from Audio Frequency Generators 1 and 2
(individually) through the Modulation Distribution assembly to the
AUDIO OUT connector, externally to the rear-panel MODULATION
IN connector, then again through the Modulation Distribution
assembly to the monitor select output and to the DVM.
The above-measured levels are used to adjust the output level of the
audio generators so that they produce a calibrated level to the
modulation inputs of the RF generator. These measurements are made
at DC. Both positive and negative levels are measured to produce an
optimum calibration factor.
SGND
+REF
-REF
Measurement Board Assembly
a2a33_1.eps
Содержание 8935 Series
Страница 34: ...34 Chapter 1 General Information Trademark Acknowledgements...
Страница 35: ...35 2 Product Information This chapter contains general information about the Test Set and how to service it...
Страница 93: ...92 Chapter 3 Troubleshooting Product Verification...
Страница 102: ...101 5 Disassembly This chapter explains how to disassemble the Test Set for major assembly replacement...
Страница 115: ...114 Chapter 5 Disassembly A2 Disassembly Figure 5 9 PC Board Assemblies...
Страница 135: ...134 Chapter 5 Disassembly Wire Cable Information...
Страница 139: ...138 Chapter 6 Replaceable Parts Parts Identification Covers and Chassis Parts Figure 6 2 External and Internal Covers...
Страница 150: ...Chapter 6 149 Replaceable Parts Parts Identification Figure 6 14 Cables Panel and Bottom Side Views...
Страница 157: ...156 Chapter 6 Replaceable Parts Parts List...
Страница 158: ...157 7 Periodic Adjustments This chapter contains the periodic adjustment procedures for the Test Set...
Страница 173: ...172 Chapter 7 Periodic Adjustments Eb No Calibration Program Description...
Страница 233: ...232 Chapter 8 Performance Tests CDMA Analyzer Modulation Accuracy Performance Test 29...
Страница 283: ...282 Chapter 9 Performance Test Records CDMA Analyzer Modulation Accuracy Performance Test 29 Record...
Страница 321: ...320 Chapter 10 Block Diagrams Instrument Control Section...
Страница 322: ...321 A Error Messages...