![Agilent Technologies 8935 Series Скачать руководство пользователя страница 224](http://html1.mh-extra.com/html/agilent-technologies/8935-series/8935-series_manual_2868625224.webp)
Chapter 8
223
Performance Tests
CDMA Generator Modulation Accuracy Performance Test 26
CDMA Generator Modulation Accuracy
Performance Test 26
The purpose of this test is to verify that the Test Set meets the
specification limits in PTR (Performance Test Record)
Table 9-26,
“CDMA Generator RF In/Out Test,” on page 277
. The modulation
accuracy of the CDMA generator is directly measured with a vector
signal analyzer at the DUPLEX OUT port. Because the vector signal
analyzer cannot measure rho directly, the modulation accuracy is
measured in EVM (Error Vector Magnitude) % rms and rho is
calculated from the EVM data.
Setup
Figure 8-31
CDMA Generator Modulation Accuracy Test 26
Procedure
1. On the Test Set:
a. Press
Preset
.
b. Press
CDMA GEN
key
c. Set
RF Gen Freq
to 836.52 MHz.
d. Set
CW RF
path to
IQ
.
e. Set
Output Port
to
Dupl
.
f. Set
Amplitude
to
−
10 dBM.
g. Set
Gen Dir
to
Rev
.
DUPLEX OUT
Vector Signal Analyzer
INPUT
Содержание 8935 Series
Страница 34: ...34 Chapter 1 General Information Trademark Acknowledgements...
Страница 35: ...35 2 Product Information This chapter contains general information about the Test Set and how to service it...
Страница 93: ...92 Chapter 3 Troubleshooting Product Verification...
Страница 102: ...101 5 Disassembly This chapter explains how to disassemble the Test Set for major assembly replacement...
Страница 115: ...114 Chapter 5 Disassembly A2 Disassembly Figure 5 9 PC Board Assemblies...
Страница 135: ...134 Chapter 5 Disassembly Wire Cable Information...
Страница 139: ...138 Chapter 6 Replaceable Parts Parts Identification Covers and Chassis Parts Figure 6 2 External and Internal Covers...
Страница 150: ...Chapter 6 149 Replaceable Parts Parts Identification Figure 6 14 Cables Panel and Bottom Side Views...
Страница 157: ...156 Chapter 6 Replaceable Parts Parts List...
Страница 158: ...157 7 Periodic Adjustments This chapter contains the periodic adjustment procedures for the Test Set...
Страница 173: ...172 Chapter 7 Periodic Adjustments Eb No Calibration Program Description...
Страница 233: ...232 Chapter 8 Performance Tests CDMA Analyzer Modulation Accuracy Performance Test 29...
Страница 283: ...282 Chapter 9 Performance Test Records CDMA Analyzer Modulation Accuracy Performance Test 29 Record...
Страница 321: ...320 Chapter 10 Block Diagrams Instrument Control Section...
Страница 322: ...321 A Error Messages...
Страница 337: ...336 Appendix A Error Messages Non Recoverable Firmware Error...