Chapter 7
169
Periodic Adjustments
Setting the Timebase Latches
Setting the Timebase Latches
The
refs_DAC_coarse
and
ref_DAC_fine
values adjust the frequency
of the Test Set’s internal 10 MHz reference. They are stored in memory.
The controller reads the values and sends the appropriate adjustment
to the A2A23 Reference assembly.
The following procedure is to be used when running the program
“Timebase Reference Using a Counter” on page 165
.
1. Press
Shift
,
Duplex Config
to access the CONFIGURE screen.
2. Select
SERVICE
under the
To Screen
menu.
3. Connect a frequency counter to the rear-panel 10 MHz REF
OUTPUT connector.
4. Select the
Latch
field.
5. Select
refs_DAC_coarse
under the
Choices:
menu.
6. Select the
Value
field.
7. Rotate the knob until the counter reads as close to 10 MHz as
possible.
8. Select the
Latch
field.
9. Select
refs_DAC_fine
under the
Choices:
menu.
10.Select the
Value
field.
11.Rotate the knob until the counter reads as close to 10 MHz as
possible.
12.Store the new DAC values (timebase calibration data) in non-volatile
memory by selecting and running the
Timebase Reference Using
a Counter
routine from the Periodic Calibration Menu. See
“Timebase Reference Using a Counter” on page 165
.
Содержание 8935 Series
Страница 34: ...34 Chapter 1 General Information Trademark Acknowledgements...
Страница 35: ...35 2 Product Information This chapter contains general information about the Test Set and how to service it...
Страница 93: ...92 Chapter 3 Troubleshooting Product Verification...
Страница 102: ...101 5 Disassembly This chapter explains how to disassemble the Test Set for major assembly replacement...
Страница 115: ...114 Chapter 5 Disassembly A2 Disassembly Figure 5 9 PC Board Assemblies...
Страница 135: ...134 Chapter 5 Disassembly Wire Cable Information...
Страница 139: ...138 Chapter 6 Replaceable Parts Parts Identification Covers and Chassis Parts Figure 6 2 External and Internal Covers...
Страница 150: ...Chapter 6 149 Replaceable Parts Parts Identification Figure 6 14 Cables Panel and Bottom Side Views...
Страница 157: ...156 Chapter 6 Replaceable Parts Parts List...
Страница 158: ...157 7 Periodic Adjustments This chapter contains the periodic adjustment procedures for the Test Set...
Страница 173: ...172 Chapter 7 Periodic Adjustments Eb No Calibration Program Description...
Страница 233: ...232 Chapter 8 Performance Tests CDMA Analyzer Modulation Accuracy Performance Test 29...
Страница 283: ...282 Chapter 9 Performance Test Records CDMA Analyzer Modulation Accuracy Performance Test 29 Record...
Страница 321: ...320 Chapter 10 Block Diagrams Instrument Control Section...
Страница 322: ...321 A Error Messages...
Страница 337: ...336 Appendix A Error Messages Non Recoverable Firmware Error...