![Agilent Technologies 8935 Series Скачать руководство пользователя страница 207](http://html1.mh-extra.com/html/agilent-technologies/8935-series/8935-series_manual_2868625207.webp)
206
Chapter 8
Performance Tests
AF Analyzer Frequency Accuracy at 400 kHz Performance Test 17
AF Analyzer Frequency Accuracy at 400 kHz
Performance Test 17
The purpose of this test is to verify that the Test Set meets the
specification limits in PTR (Performance Test Record)
Table 9-17, “AF
Analyzer Frequency Accuracy at 400 kHz Test 17 Record,” on page 267
.
To measure frequency accuracy at 400 kHz, the RF signal from the Test
Set’s DUPLEX OUT port is applied to the audio input and the input to
the measuring receiver and the two measured frequencies are
compared.
Figure 8-22
Setup
Procedure
1. On the measuring receiver:
a. Reset the instrument.
b. Set the measurement mode to frequency.
2. On the Test Set:
a. Press
Preset
.
b. Select the INSTRUMENT CONFIGURE screen.
c. Set the
RF Display
field to
Freq
.
d. Select the CDMA GENERATOR screen.
e. Set the
CW RF Path
field to
Bypass
.
f. Select the RF GENERATOR screen.
g. Set the
RF Gen Freq
to
0.4 MHz
.
h. Set the
Amplitude
to
-21 dBm
(20 mV).
i. Select the AF ANALYZER screen.
j. Set the
AF Anl In
field to Audio In.
k. Set the
Filter 2
field to
>99kHz LPF
.
AUDIO IN HI
DUPLEX OUT
INPUT 50
Ω
Measuring Receiver
Содержание 8935 Series
Страница 34: ...34 Chapter 1 General Information Trademark Acknowledgements...
Страница 35: ...35 2 Product Information This chapter contains general information about the Test Set and how to service it...
Страница 93: ...92 Chapter 3 Troubleshooting Product Verification...
Страница 102: ...101 5 Disassembly This chapter explains how to disassemble the Test Set for major assembly replacement...
Страница 115: ...114 Chapter 5 Disassembly A2 Disassembly Figure 5 9 PC Board Assemblies...
Страница 135: ...134 Chapter 5 Disassembly Wire Cable Information...
Страница 139: ...138 Chapter 6 Replaceable Parts Parts Identification Covers and Chassis Parts Figure 6 2 External and Internal Covers...
Страница 150: ...Chapter 6 149 Replaceable Parts Parts Identification Figure 6 14 Cables Panel and Bottom Side Views...
Страница 157: ...156 Chapter 6 Replaceable Parts Parts List...
Страница 158: ...157 7 Periodic Adjustments This chapter contains the periodic adjustment procedures for the Test Set...
Страница 173: ...172 Chapter 7 Periodic Adjustments Eb No Calibration Program Description...
Страница 233: ...232 Chapter 8 Performance Tests CDMA Analyzer Modulation Accuracy Performance Test 29...
Страница 283: ...282 Chapter 9 Performance Test Records CDMA Analyzer Modulation Accuracy Performance Test 29 Record...
Страница 321: ...320 Chapter 10 Block Diagrams Instrument Control Section...
Страница 322: ...321 A Error Messages...
Страница 337: ...336 Appendix A Error Messages Non Recoverable Firmware Error...