Chapter 7
171
Periodic Adjustments
Eb/No Calibration Program Description
Eb/No Calibration Program Description
The Eb/No calibration is a CDMA loopback measurement. Before the
Eb/No measurement begins, four preliminary measurements are made
with the CDMA generator in the data mode: (1) The difference in
channel power between forward and reverse modes is measured to
determine the loss to be accounted for when the all-pass filter is in
forward mode. (2) Rho is measured in both the forward and reverse
paths. (3) The time offset is measured and its value entered. (4) Fast
power is measured to assure that the signal is “noiseless.”
The generator is then set to the Eb/No mode at values of 7, 10, 12, 14,
16, 18, and 20 dB, and one hundred measurements of fast power are
made at each value. (This takes several minutes to complete.) The
measured data is then processed and converted into calibration factors.
Figure 7-6
Eb/No Calibration
servsrn1.eps
SOFTWARE MENU Screen
SOFTWARE MENU
Description:
Launches disgnostic and calibration programs.
Continue
1
2
4
Run Test
Help
SOFTWARE MENU
Library:
[NO LIB]
Program:
ROM
CUSTOMIZE TEST PROCEDURE:
Channel Information
Test Parameters
Order of Tests
Pass/Fail Limits
Save/Delete Procedure
SET UP TEST SET:
Execution Cond
External Devices
Printer Setup
IBASIC Cntrl
LOAD TEST PROCEDURE:
Select Procedure Location:
ROM
Select Procedure Filename:
SERVICE4
Exec
Cnfg
Print
IBASIC
Freq
Parm
Seqn
Spec
Proc
ebno1.eps
SERVICE MENU
1
2
3
4
5
Help
Exit
Functional Diagnostics
AF Diagnostics
RF Diagnostics
CDMA Diagnostics
Edit RF Diagnostic Limits
Periodic Calibration
IQ Calibration
Eb/No Calibration
=>
Move pointer to the desired program using the
knob then press the knob. Press HELP for
information on the tests. Press EXIT to abort.
SERVICE MENU Screen
Содержание 8935 Series
Страница 34: ...34 Chapter 1 General Information Trademark Acknowledgements...
Страница 35: ...35 2 Product Information This chapter contains general information about the Test Set and how to service it...
Страница 93: ...92 Chapter 3 Troubleshooting Product Verification...
Страница 102: ...101 5 Disassembly This chapter explains how to disassemble the Test Set for major assembly replacement...
Страница 115: ...114 Chapter 5 Disassembly A2 Disassembly Figure 5 9 PC Board Assemblies...
Страница 135: ...134 Chapter 5 Disassembly Wire Cable Information...
Страница 139: ...138 Chapter 6 Replaceable Parts Parts Identification Covers and Chassis Parts Figure 6 2 External and Internal Covers...
Страница 150: ...Chapter 6 149 Replaceable Parts Parts Identification Figure 6 14 Cables Panel and Bottom Side Views...
Страница 157: ...156 Chapter 6 Replaceable Parts Parts List...
Страница 158: ...157 7 Periodic Adjustments This chapter contains the periodic adjustment procedures for the Test Set...
Страница 173: ...172 Chapter 7 Periodic Adjustments Eb No Calibration Program Description...
Страница 233: ...232 Chapter 8 Performance Tests CDMA Analyzer Modulation Accuracy Performance Test 29...
Страница 283: ...282 Chapter 9 Performance Test Records CDMA Analyzer Modulation Accuracy Performance Test 29 Record...
Страница 321: ...320 Chapter 10 Block Diagrams Instrument Control Section...
Страница 322: ...321 A Error Messages...
Страница 337: ...336 Appendix A Error Messages Non Recoverable Firmware Error...