14.Enter the firmware code word: This code word is unit specific and
can only be used with the unit for which it was ordered. If you have
more than one test set to upgrade you must associate the upgrade kit
with the serial number of the test set for which it was ordered.
15.Once the codeword is entered the program will prompt you to cycle
power to enable the CDMA 2000 feature.
After the Test set has fully booted you must choose the IS-2000 function
by selecting the Inst Config screen:
Move the cursor to the CDMA Std field on the Instrument Configure
screen and select IS-2000 to enable the CDMA 2000 functionality.
If you encounter any difficulty with this upgrade please email or call
Technical Support for additional assistance.
Checking Firmware Version
The following procedure checks the current version of firmware in the
Test Set.
1. Power on the Test Set.
2. Press the
Inst Config
key. The INSTRUMENT CONFIGURE screen
appears. The host firmware level appears in the upper right corner
of this screen.
3. To find the Receive (RX) DSP firmware revision level, select the title
bar of the INSTRUMENT CONFIGURE screen. A drop down menu
appears.
4. Select SERVICE. The SERVICE screen appears, see
Figure 2-3 on
page 43
.
Содержание 8935 Series
Страница 34: ...34 Chapter 1 General Information Trademark Acknowledgements...
Страница 35: ...35 2 Product Information This chapter contains general information about the Test Set and how to service it...
Страница 93: ...92 Chapter 3 Troubleshooting Product Verification...
Страница 102: ...101 5 Disassembly This chapter explains how to disassemble the Test Set for major assembly replacement...
Страница 115: ...114 Chapter 5 Disassembly A2 Disassembly Figure 5 9 PC Board Assemblies...
Страница 135: ...134 Chapter 5 Disassembly Wire Cable Information...
Страница 139: ...138 Chapter 6 Replaceable Parts Parts Identification Covers and Chassis Parts Figure 6 2 External and Internal Covers...
Страница 150: ...Chapter 6 149 Replaceable Parts Parts Identification Figure 6 14 Cables Panel and Bottom Side Views...
Страница 157: ...156 Chapter 6 Replaceable Parts Parts List...
Страница 158: ...157 7 Periodic Adjustments This chapter contains the periodic adjustment procedures for the Test Set...
Страница 173: ...172 Chapter 7 Periodic Adjustments Eb No Calibration Program Description...
Страница 233: ...232 Chapter 8 Performance Tests CDMA Analyzer Modulation Accuracy Performance Test 29...
Страница 283: ...282 Chapter 9 Performance Test Records CDMA Analyzer Modulation Accuracy Performance Test 29 Record...
Страница 321: ...320 Chapter 10 Block Diagrams Instrument Control Section...
Страница 322: ...321 A Error Messages...
Страница 337: ...336 Appendix A Error Messages Non Recoverable Firmware Error...