Chapter 3
51
Troubleshooting
Self-Test Diagnostics (Step 1)
To Start Troubleshooting
1. Turn on the Test Set to automatically run the self test diagnostics.
• If the Test Set does not power up, see
“If the Test Set Fails to
Power-up” on page 52
.
• If all self-test diagnostics pass, and the front-panel keys and knob
work, you can assume that the digital control assemblies work.
2. After power-up, the top line of the Test Set’s display should show
copyright information and the firmware revision code. The second
line should display
All self tests passed
.
• If the Test Set powers-up with
“One or more self-tests
failed. Error code:<hexadecimal error code>:”
, see
“Reading Front Panel or GPIB Codes” on page 54
.
• See
“Frequently Encountered Diagnostic Messages” on page 73
.
for other error messages that might appear on the second line of
the display.
3. The CDMA ANALYZER screen should be displayed. Two conditions
cause a different screen to be displayed on power-up:
❏
A SAVE/RECALL register named POWERON was saved to
automatically power-up the Test Set in a different state. Press the
Preset
key before proceeding; this will restore the Test Set to the
factory power-up condition.
❏
The
Autostart Test Procedure on Power-Up:
field (of the
“TESTS [Execution Conditions]” screen) is set to
On
to
automatically run a loaded program. Press the
Shift
key, then
press the
Cancel
key to stop the program. Press the
Preset
key to
restore the Test Set to the factory power-up condition.
To turn the autostart function off, press the
Menu
key, then select
Execution Cond
(under the
SET UP TEST SET:
heading). The
autostart function is at the bottom of the screen; turn it
Off
.
Содержание 8935 Series
Страница 34: ...34 Chapter 1 General Information Trademark Acknowledgements...
Страница 35: ...35 2 Product Information This chapter contains general information about the Test Set and how to service it...
Страница 93: ...92 Chapter 3 Troubleshooting Product Verification...
Страница 102: ...101 5 Disassembly This chapter explains how to disassemble the Test Set for major assembly replacement...
Страница 115: ...114 Chapter 5 Disassembly A2 Disassembly Figure 5 9 PC Board Assemblies...
Страница 135: ...134 Chapter 5 Disassembly Wire Cable Information...
Страница 139: ...138 Chapter 6 Replaceable Parts Parts Identification Covers and Chassis Parts Figure 6 2 External and Internal Covers...
Страница 150: ...Chapter 6 149 Replaceable Parts Parts Identification Figure 6 14 Cables Panel and Bottom Side Views...
Страница 157: ...156 Chapter 6 Replaceable Parts Parts List...
Страница 158: ...157 7 Periodic Adjustments This chapter contains the periodic adjustment procedures for the Test Set...
Страница 173: ...172 Chapter 7 Periodic Adjustments Eb No Calibration Program Description...
Страница 233: ...232 Chapter 8 Performance Tests CDMA Analyzer Modulation Accuracy Performance Test 29...
Страница 283: ...282 Chapter 9 Performance Test Records CDMA Analyzer Modulation Accuracy Performance Test 29 Record...
Страница 321: ...320 Chapter 10 Block Diagrams Instrument Control Section...
Страница 322: ...321 A Error Messages...
Страница 337: ...336 Appendix A Error Messages Non Recoverable Firmware Error...