Chapter 3
73
Troubleshooting
Frequently Encountered Diagnostic Messages
Frequently Encountered Diagnostic Messages
Warning/Error Messages
Error messages that appear on the second line of the Test Set’s display
frequently occur while any of the SERVICE4 program diagnostic tests
are running. The most complete and general list of error messages is in
the
“Error Messages
” chapter of the Test Set’s Reference Guide. (Some
messages relating specifically to troubleshooting can be found in
Appendix A , “Error Messages,” on page 321
.) Some of the messages you
can expect to occur while running the SERVICE4 program diagnostic
tests are as follows:
•
Direct latch write occurred. Cycle power when done
servicing.
The SERVICE4 program commonly generates this
message. This message appears the first time the diagnostic
program directly addresses a latch. The message should be ignored
and cleared when you make a normal (not a diagnostic)
measurement with the Test Set. To clear this message the Test Set
should be turned off and back on again.
•
Change Ref Level, Input Port or Attenuator (if using
“Hold”).
This message, and similar messages, can be generally
ignored.
•
Printer does not respond.
This usually indicates that one or
more settings on the TESTS (Printer Setup) screen are set
incorrectly for your printer. Also, check that the printer’s power is on
and that it is correctly cabled. For GPIB printers make sure the
printer is correctly addressed. If a serial printer is used, you may
have to change the serial communication settings on the I/O
CONFIGURE screen (press
Shift
then
Inst Config
to get to this
screen). The message times out after a few seconds, and the output
destination is changed to CRT by the program.
•
ERROR 173 IN XXXX Active/system controller req’d
(where
”XXXX”
represents a line number). Indicates that the Test Set’s
internal IBASIC computer must be set as a system controller for
some reason. This usually indicates that the
Printer Port
field of
the TESTS (Printer Setup) screen was set to
HP-IB
but the
Mode
field
on the I/O CONFIGURE screen is set to
Talk&Lstn
instead of
Control
. Change the mode setting to
Control
and run the
diagnostic again.
Содержание 8935 Series
Страница 34: ...34 Chapter 1 General Information Trademark Acknowledgements...
Страница 35: ...35 2 Product Information This chapter contains general information about the Test Set and how to service it...
Страница 93: ...92 Chapter 3 Troubleshooting Product Verification...
Страница 102: ...101 5 Disassembly This chapter explains how to disassemble the Test Set for major assembly replacement...
Страница 115: ...114 Chapter 5 Disassembly A2 Disassembly Figure 5 9 PC Board Assemblies...
Страница 135: ...134 Chapter 5 Disassembly Wire Cable Information...
Страница 139: ...138 Chapter 6 Replaceable Parts Parts Identification Covers and Chassis Parts Figure 6 2 External and Internal Covers...
Страница 150: ...Chapter 6 149 Replaceable Parts Parts Identification Figure 6 14 Cables Panel and Bottom Side Views...
Страница 157: ...156 Chapter 6 Replaceable Parts Parts List...
Страница 158: ...157 7 Periodic Adjustments This chapter contains the periodic adjustment procedures for the Test Set...
Страница 173: ...172 Chapter 7 Periodic Adjustments Eb No Calibration Program Description...
Страница 233: ...232 Chapter 8 Performance Tests CDMA Analyzer Modulation Accuracy Performance Test 29...
Страница 283: ...282 Chapter 9 Performance Test Records CDMA Analyzer Modulation Accuracy Performance Test 29 Record...
Страница 321: ...320 Chapter 10 Block Diagrams Instrument Control Section...
Страница 322: ...321 A Error Messages...
Страница 337: ...336 Appendix A Error Messages Non Recoverable Firmware Error...