![Agilent Technologies 8935 Series Скачать руководство пользователя страница 104](http://html1.mh-extra.com/html/agilent-technologies/8935-series/8935-series_manual_2868625104.webp)
Chapter 5
103
Disassembly
Assembly Replacements
Assembly Replacements
With some assemblies you will receive a memory card that contains
factory-generated calibration data for that assembly. For new
replacements, there will also be an instruction sheet for loading the
calibration data into your Test Set.
External equipment is not required for running the diagnostic routines.
If diagnostic routines cannot isolate the problem, an oscilloscope,
voltmeter, and spectrum analyzer may be required for further
troubleshooting. A second Test Set is helpful for troubleshooting
performance test failures.
Table 3-2 on page 75
and
Table 7-1 on page 159
show which assemblies
need calibration data as well as which performance tests and periodic
self-calibration adjustments are recommended after replacing an
assembly.
NOTE
Periodic Adjustment Intervals
The adjustment programs Periodic Calibration, IQ Calibration, and
Eb/No Calibration should be performed after any assembly referred to
in
Table 7-1 on page 159
is replaced, or at least every 24 months. These
program can be run anytime to optimize the performance of the Test
Set. See
Chapter 7 , “Periodic Adjustments,” on page 157
for details.
NOTE
Performance Test Intervals
The performance tests in
Chapter 8 , “Performance Tests,” on page 173
should be performed when certain assemblies are repaired or replaced,
or at least every 24 months. See
Table 3-2 on page 75
for those
assemblies requiring performance testing/calibration.
Replacement Parts
For replacement part numbers, see
Chapter 6 , “Replaceable Parts,” on
page 135
. For cable routing information refer to
Table 5-2 on page 130
.
Содержание 8935 Series
Страница 34: ...34 Chapter 1 General Information Trademark Acknowledgements...
Страница 35: ...35 2 Product Information This chapter contains general information about the Test Set and how to service it...
Страница 93: ...92 Chapter 3 Troubleshooting Product Verification...
Страница 102: ...101 5 Disassembly This chapter explains how to disassemble the Test Set for major assembly replacement...
Страница 115: ...114 Chapter 5 Disassembly A2 Disassembly Figure 5 9 PC Board Assemblies...
Страница 135: ...134 Chapter 5 Disassembly Wire Cable Information...
Страница 139: ...138 Chapter 6 Replaceable Parts Parts Identification Covers and Chassis Parts Figure 6 2 External and Internal Covers...
Страница 150: ...Chapter 6 149 Replaceable Parts Parts Identification Figure 6 14 Cables Panel and Bottom Side Views...
Страница 157: ...156 Chapter 6 Replaceable Parts Parts List...
Страница 158: ...157 7 Periodic Adjustments This chapter contains the periodic adjustment procedures for the Test Set...
Страница 173: ...172 Chapter 7 Periodic Adjustments Eb No Calibration Program Description...
Страница 233: ...232 Chapter 8 Performance Tests CDMA Analyzer Modulation Accuracy Performance Test 29...
Страница 283: ...282 Chapter 9 Performance Test Records CDMA Analyzer Modulation Accuracy Performance Test 29 Record...
Страница 321: ...320 Chapter 10 Block Diagrams Instrument Control Section...
Страница 322: ...321 A Error Messages...
Страница 337: ...336 Appendix A Error Messages Non Recoverable Firmware Error...