General Maintenance
4
Troubleshooting and Maintenance Manual
99
To drain/refill the foreline pump
Materials needed
•
Book or other solid object approximately 5 cm thick
•
Container for catching old pump oil, 500 mL
•
Gloves, oil- and solvent-resistant
•
Screwdriver, flat-blade, large (8730-0002)
•
Hex key, 5-mm (8710-1838)
•
Foreline pump oil (6040-0834)
Drain the pump
1
Vent the MS. See the
Agilent 7000 Series Triple Quad GC/MS Operation
.
2
If necessary, slide the foreline pump to a safe, accessible location.
The foreline pump may be located on the floor, on the lab bench next to or
behind the MS, or under the analyzer chamber at the back of the MS.
3
Place a book or other object under the back of the pump motor to tilt the
front of it down slightly. Remove the fill cap. See
4
Place a container under the drain plug.
5
Remove the drain plug. Allow the pump oil to drain out. The oil drains
faster if it is still warm.
WA R N I N G
The foreline pump can cause burns if touched when operating. An optional safety
shield will prevent the user from touching it.
WA R N I N G
The foreline pump may be hot.
WA R N I N G
The old pump oil may contain toxic chemicals. Treat it as hazardous waste.
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...