40
Troubleshooting and Maintenance Manual
2
General Troubleshooting
Temperature Symptoms
The MS has four heated zones:
•
Ion source
•
Front and rear mass filters
•
GC/MS interface
Each heated zone has a heater and temperature sensor. The ion source and
mass filters are powered and controlled by the MS. The GC/MS interface is
powered and controlled by the GC.
Ion source will not heat up
•
High-vacuum pump is off or has not reached normal operating conditions*
•
Incorrect temperature setpoint
•
Ion source has not had enough time to reach temperature setpoint
•
Ion source heater cartridge is not connected*
•
Ion source temperature sensor is not connected*
•
Ion source heater failed (burned out or shorted to ground)*
•
Ion source temperature sensor failed*
•
Source power cable is not connected to the quadrupole board*
•
MS electronics are not working correctly
* This will cause an error message.
Mass filter (quad) heaters will not heat up
•
High-vacuum pump is off or has not reached normal operating conditions*
•
Incorrect temperature setpoint
•
Mass filter has not had enough time to reach temperature setpoint
•
Mass filter heater cartridge is not connected*
•
Mass filter temperature sensor is not connected*
•
Mass filter heater failed (burned out or shorted to ground)*
•
Mass filter temperature sensor failed*
•
Cable is not connected to the quadrupole board*
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...