
46
Troubleshooting and Maintenance Manual
2
General Troubleshooting
There is no emission current
•
Check tune file to be certain that emission current is not = 0
•
Filament is not connected properly; try the other filament
•
Filament has failed; try the other filament
•
MS electronics are not working correctly
There is not enough signal to begin tune
•
Corrupted tune file
•
Poor mass axis calibration
•
Width gain or offset is too high
•
Calibration vial(s) empty or almost empty
•
Excessive pressure in the analyzer chamber
•
Air leak
•
Electron multiplier voltage is too low
•
Signal cable is not connected
•
Electrical leads to the detector are not connected correctly
•
Electrical leads to the ion source are not connected correctly
•
Filament shorted to the source body
•
Collision cell gas flow
•
Collision cell voltages
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...