74
Troubleshooting and Maintenance Manual
3
CI Troubleshooting
Peak widths are unstable
Wrong or corrupted tune file
Check the tune parameters.
The CI ion source is dirty.
on page 123 for more
information.
Air leak
Run Methane Pretune (in PCI). A visible peak at
m/z
32 indicates air in the
system. Check for and correct any leaks. See
eliminating all air leaks, clean the ion source.