58
Troubleshooting and Maintenance Manual
3
CI Troubleshooting
Pressure-Related Symptoms
The following symptoms are all related to high vacuum pressure. Each
symptom is discussed in more detail in the following pages.
The mass flow controller is calibrated for methane and the high vacuum gauge
controller is calibrated for nitrogen, so these measurements are not accurate
in absolute terms (
). They are intended as a guide to typical observed
readings. They were taken with the following set of conditions:
Source temperature
300 °C
Quad temperature
150 °C
Interface temperature
280 °C to 320 °C
Helium carrier gas flow
1 mL/min
Table 5
Typical analyzer vacuum with reagent gas flow
Collision cell gas flow on
Collision cell gas flow off
MFC
(%)
Rough vac
High vac
Rough vac
High vac
10
1.77
×
10
–1
7.15
×
10
–5
1.33
×
10
–1
2.56
×
10
–6
15
1.86
×
10
–1
7.19
×
10
–5
1.43
×
10
–1
3.00
×
10
–6
20
1.94
×
10
–1
7.23
×
10
–5
1.53
×
10
–1
3.45
×
10
–6
25
2.02
×
10
–1
7.27
×
10
–5
1.63
×
10
–1
3.86
×
10
–6
30
2.10
×
1
0
–1
7.31
×
10
–5
1.71
×
1
0
–1
4.30
×
10
–6
35
2.18
×
10
–1
7.39
×
10
–5
1.80
×
10
–1
4.76
×
10
–6
40
2.25
×
10
–1
7.43
×
10
–5
1.88
×
10
–1
5.18
×
10
–6
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...