General Maintenance
4
Troubleshooting and Maintenance Manual
113
Reinstallation
1
Unpack the new mass filter heater assembly (
). The heater,
temperature sensor, and heater block are already assembled.
2
Install the heater assembly on top of the mass filter radiator using the two
screws. See
for replacing the front mass filter heater/sensor. The
rear mass filter heater sensor is similar.
3
Connect the heater and temperature sensor wires to the feedthrough
board.
4
Close the analyzer chamber.
5
Pump down the MS. See the
Agilent 7000 Series Triple Quad GC/MS
.
C A U T I O N
Do not touch the mass filter contact leads. This could cause ESD damage to the
quadrupole board.
C A U T I O N
Do not touch the mass filter contact leads. This could cause ESD damage to the
quadrupole board.