16
Troubleshooting and Maintenance Manual
1
Introduction
•
Accumulation of hydrogen in the GC oven and subsequent combustion (See
your GC documentation and the label on the top edge of the GC oven door.)
•
Accumulation of hydrogen in the MS and subsequent combustion
Hydrogen accumulation in an MS
All users should be aware of the mechanisms by which hydrogen can
accumulate (
) and know what precautions to take if they are certain or
suspect that hydrogen has accumulated. Note that these mechanisms apply to
all
mass spectrometers.
WA R N I N G
The MS cannot detect leaks in inlet and/or detector gas streams. For this reason, it
is vital that column fittings should always be either connected to a column or have a
cap or plug installed.
Table 2
Hydrogen accumulation mechanisms
Mechanism
Results
Mass spectrometer turned off
A mass spectrometer can be shut down
deliberately. It can also be shut down accidentally
by an internal or external failure. A mass
spectrometer shutdown does not shut off the flow
of carrier gas. As a result, hydrogen may slowly
accumulate in the mass spectrometer.
Mass spectrometer automated shutoff
valves closed
The mass spectrometers are equipped with
automated shutoff valves for the calibration vial
and the reagent gases. Deliberate operator action
or various failures can cause the shutoff valves to
close. Shutoff valve closure does not shut off the
flow of carrier gas. As a result, hydrogen may
slowly accumulate in the mass spectrometer.
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...