General Maintenance
4
Troubleshooting and Maintenance Manual
87
6
Place a column nut with a blank ferrule on the end of the interface. This
will help keep contamination out of the MS.
7
Disconnect the GC/MS interface cable. Disconnecting the cable with the
GC on can cause a fault condition.
8
Continue to move the MS until you have access to the part requiring
maintenance.
Reconnection
1
Position the MS so the end of the GC/MS interface is near the GC
(
2
Reconnect the GC/MS interface cable.
Figure 6
Separating/connecting the MS and GC
GC/MS interface
Interface cable