General Maintenance
1
Slide the new heater and temperature sensor into the heater sleeve
(
2
Reinstall the setscrew.
3
Slide the heater sleeve onto the GC/MS interface. Align the heater sleeve
so the screws are on top. Tighten the screws evenly.
4
Slide the insulation onto the GC/MS interface.
5
Reinstall the GC/MS interface cover. Make sure the wires from the heater
and sensor pass through the cutout in the interface cover.
6
Reconnect the MS to the GC. See
7
Reconnect the GC/MS interface cable to the GC. Reinstall the capillary
column.
8
Pump down the MS. See the
Agilent 7000 Series Triple Quad GC/MS
9
Turn on the GC. Enter appropriate temperature setpoints for the GC/MS
interface and GC oven.
C A U T I O N
There is a shallow groove along the inner surface of the insulation. This groove
must
line up with the heads of the screws in the heater sleeve. If it does not, you can crack
or otherwise damage the insulation.
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...