Troubleshooting and Maintenance Manual
2
General Troubleshooting
•
Calibration valve is not working correctly
•
Bad signal cable connection
•
Filament has failed or is not connected correctly
•
Bad ion source wiring connection
•
Bad detector wiring connection
•
Failed electron multiplier horn
Peaks are tailing
•
Active sites in the sample path
•
Injection is too large
•
Incorrect GC inlet temperature
•
Insufficient column flow
•
GC/MS interface temperature is too low
•
Ion source temperature is too low
Peaks are fronting
•
Column film thickness mismatched with analyte concentration (column
overload)
•
Initial oven temperature is too low
•
Active sites in the sample path
•
Injection is too large
•
GC inlet pressure too high
•
Insufficient column flow
Peaks have flat tops
•
Insufficient solvent delay
•
Incorrect scale on the display
•
Injection is too large
•
Electron multiplier voltage is too high
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...