Troubleshooting and Maintenance Manual
4
General Maintenance
9
Turn the collar clockwise to tighten it.
The collar should be snug but not overly tight. Do
not
use a tool to tighten
the collar. It does not require that much force.
10
Reinstall the analyzer window cover.
11
In
Instrument Control
panel, select the
MS Tune
icon to display the
GC-QQQ
Tune
dialog box. Select the
Manual Tune
tab then select the
Ion Source
tab to
display the ion source parameters.
12
Turn off the
Emission
by clicking in the check box.
13
Purge the calibration valve by clicking in the
EI Cal Valve
check box to open
the calibration valve. Close the
EI Cal Valve
after 30 seconds.
C A U T I O N
After removing a calibrant vial, you
must
purge the calibration valve. Failure to do so
will result in damage to the filaments and the electron multiplier.
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...