General Maintenance
4
Troubleshooting and Maintenance Manual
101
To replace the fan for the turbo pump
Procedure
1
Vent the MS. See the
Agilent 7000 Series Triple Quad GC/MS Operation
.
2
Turn the MS off.
3
Move the MS away from the GC to make the fan accessible.
4
Remove the fan plenum (four screws).
5
Disconnect the fan wiring from the extender cable (
).
6
Remove the four fan screws and the safety grill. Remove the fan. Keep the
screws.
7
Connect the fan wiring to the new fan.
8
Install the new fan with the flow arrow on the side pointing toward the
pump.
9
Add the safety grill and the four screws. Tighten the screws firmly.
10
Connect the fan wiring to the extender cable.
11
Pump down the MS. See the
Agilent 7000 Series Triple Quad GC/MS
WA R N I N G
Do not touch the high vacuum pump. The pump could still be hot enough to burn you.
WA R N I N G
Make sure the safety grill that shields the fan blades is in place.
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...