General Troubleshooting
2
Troubleshooting and Maintenance Manual
29
Chromatographic Symptoms
These are symptoms you may observe in the chromatograms generated by data
acquisition. In general, these symptoms do not prevent you from operating
your GC/MS system. They indicate, however, that the data you are acquiring
may not be the best data obtainable. These symptoms can be caused by
instrument malfunctions but are more likely caused by incorrect
chromatographic technique.
Two of the symptoms,
Low sensitivity
and
Poor repeatability
also apply to
mass spectral data.
No peaks
If an analysis shows no chromatographic peaks, only a flat baseline or minor
noise, run the automated tune program. If the MS passes tune, the problem is
most likely related to the GC. If the MS does not pass tune, the problem is most
likely in the MS.
Passes tune
•
Incorrect sample concentration
•
No analytes present
•
Syringe missing from the ALS or not installed correctly
•
Injection accidentally made in split mode instead of splitless mode
•
Empty or almost empty sample vial
•
Dirty GC inlet
•
Leaking GC inlet*
•
Loose column nut at the GC inlet*
* This could cause a fault condition in the GC that would prevent the GC
from operating.
Does not pass tune
•
Calibration vial is empty
•
Excessive foreline or analyzer chamber pressure
•
Very dirty ion source
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...