General Troubleshooting
2
Troubleshooting and Maintenance Manual
39
Foreline pressure is too low
If the pressures you observe are below 20 mTorr, check for the following:
•
Column (carrier gas) flow is too low
•
Column plugged or crushed by an overtightened nut
•
Collision gas flows are too low
•
Empty or insufficient carrier gas supply*
•
Bent or pinched carrier gas tubing*
•
Foreline gauge is not working correctly
* This could create a fault condition in the GC that would prevent the GC
from operating.
Analyzer chamber pressure is too low
If the pressures you observe are below 1 × 10
-6
Torr, check for the following:
•
Column (carrier gas) flow is too low
•
Collision gas flows are too low
•
Column plugged or crushed by overtightened nut
•
Empty or insufficient carrier gas supply*
•
Bent or pinched carrier gas tubing*
* This could create a fault condition in the GC that would prevent the GC
from operating.
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...