86
Troubleshooting and Maintenance Manual
4
General Maintenance
To separate/connect the MS and the GC
Materials needed
•
Wrench, open-end, 1/4-inch × 5/16-inch (8710-0510)
Separation
1
Vent the MS. See the
Agilent 7000 Series Triple Quad GC/MS Operation
.
2
Turn off the GC.
3
Remove the capillary column from the GC/MS interface.
4
The foreline pump may be located on the floor, on the lab bench next to or
behind the MS, or under the analyzer chamber at the back of the MS. Move
it as needed to provide slack in the tubing and cables.
5
Move the MS away from the GC until you have access to the GC/MS
interface cable (
WA R N I N G
Make sure the GC/MS interface and the analyzer zones are cool (below 100 °C)
before you vent the MS. A temperature of 100 °C is hot enough to burn skin; always
wear cloth gloves when handling analyzer parts.
WA R N I N G
If you are using hydrogen as a carrier gas, the carrier gas flow must be off before
turning off the MS power. If the foreline pump is off, hydrogen will accumulate in the
MS and an explosion may occur. Read
with hydrogen carrier gas.
C A U T I O N
Be sure the GC oven and the GC/MS interface are cool before turning off carrier gas
flow.
WA R N I N G
Make sure the GC/MS interface and GC oven have cooled before you remove the
column.
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...