Excessive pressure in the front or rear chamber
•
Dirty ion source
•
Air leaks between chambers
•
Poor filament operation
•
Detector (HED electron multiplier) is not working correctly
•
Incorrect mass filter polarity
•
Collision cell voltage
* This could cause a fault condition in the GC that would prevent the GC
from operating.
Poor repeatability
•
Dirty syringe needle
•
Dirty GC inlet
•
Leaking GC inlet*
•
Injection is too large
•
Loose column connections
•
Variations in pressure, column flow, and temperature
•
Dirty ion source
•
Loose connections in the analyzer
•
Ground loops
* This could cause a fault condition in the GC that would prevent the GC
from operating.